Liquid chromatographic estimation of (S) - glycidyl butyrate in (R) - glycidyl butyrate


Analytical Research Laboratory, Glenmark Research Center, Glenmark Generics Limited, Navi Mumbai, Maharashtra, India

Date of Web Publication27-Jan-2012

Correspondence Address:
Mubeen Ahmad Khan
Senior Vice President, Glenmark Generics Limited, Glenmark Research Center, Navi Mumbai- 400 709
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0976-9668.92331

Rights and Permissions
   Abstract 

Aim: The present study was undertaken out of a commercial need for the synthesis of Linezolid with impurity limits within the specification. Materials and Methods: ( R)-Glycidyl butyrate (RGB) is raw material for the synthesis of Linezolid drug substance. This RGB contains (S)-(+)-Glycidyl butyrate (SGB) and SGB appears in same concentration in the final Active Pharmaceutical Ingredient. So, a normal phase high-performance liquid chromatography (HPLC) method has been developed to determine the SGB level in the raw material. RGB and SGB were separated using an HPLC system equipped with quaternary gradient pumps on a Daicel chiralpak AD-H (250 × 4.6 mm) column with a mobile phase consisting 2.0 ml of ethanol in 1 000 ml of n-hexane. A 0.5 ml/minute flow rate and a 10 μl injection volume was used and the compounds were detected at 215 nm. Results: Method validation parameters demonstrated the same to be reliable, reproducible, and accurate one. Conclusion: Thus, the present study may be used for regular quality control of RGB to improve commercial feasibility for the synthesis of Linezolid.

Keywords: (R)--Glycidyl butyrate, (S)--(+)--glycidyl butyrate, chiral--HPLC, Linezolid, validation


How to cite this article:
Khan MA, Sinha S, Khandekar N, Parashar V. Liquid chromatographic estimation of (S) - glycidyl butyrate in (R) - glycidyl butyrate. J Nat Sc Biol Med 2011;2:180-4

How to cite this URL:
Khan MA, Sinha S, Khandekar N, Parashar V. Liquid chromatographic estimation of (S) - glycidyl butyrate in (R) - glycidyl butyrate. J Nat Sc Biol Med [serial online] 2011 [cited 2020 Sep 24];2:180-4. Available from: http://www.jnsbm.org/text.asp?2011/2/2/180/92331


   Introduction Top


The eventual purpose of any analytical method in general and liquid chromatographic method in particular is to separate and quantify the compounds of interest. These compounds may either be drug substances or an unwanted substance in drug substance, its raw material, or intermediates. [1] The developed method should be a reproducible, accurate, sensitive, and commercially viable one. To ensure these parameters, validation of the method is performed as per guidelines like International Conference on Harmonization (ICH). [2]

Optically pure 2,3-epoxy-1-propanol (glycidol) and its derivatives are versatile intermediates in organic synthesis because the epoxide ring is reactive toward nucleophiles for the synthesis of asymmetric alcohols. [3],[4],[5] Particularly, both enantiomers of the (R)-glycidyl butyrate (RGB) and (S)-glycidyl butyrate (SGB) [Figure 1] are chiral synthesis units for the production of biologically active compounds of commercial interest. RGB has been used to introduce a stereogenic center in the synthesis of Linezolid, [6],[7] which is currently marketed for the treatment of multidrug-resistant Gram-positive infections such as nosocomial, community-acquired pneumonia, and skin infections. SGB is also very useful as starting material in the synthesis of many drugs, such as (R)-argentilactone, [8] which exhibits both antileishmanial activity and cytotoxic activity against mouse leukemia cells. [9]
Figure 1: Chemical structure of SGB and RGB

Click here to view


RGB is an intermediate in the synthesis of Linezolid drug substance. The importance of the study lies in the fact that SGB is inherently present in RGB and it appears in almost same percentage in the Linezolid drug substance as was present in the intermediate (RGB). Thus, the content of SGB, if not controlled in the raw material itself, will appear in more than specified limit in the finished drug substance (Linezolid). Literature survey on this revealed absence of any such reports or publications. Thus, the current study was undertaken out of the need for successful commercialization of Linezolid. A chiral high-performance liquid chromatography (HPLC) method has been developed for the estimation of SGB in RGB sample. [10],[11]


   Experimental Top


Reagents and chemicals

Standard SGB was obtained from Sigma Aldrich, United States of America (USA). RGB (Glenmark Generics Ltd, Glenmark Research Centre, Navi Mumbai, Maharashtra, India) was obtained as in-house standard. n-Hexane (HPLC Grade) was purchased from Ranbaxy Fine Chemicals Limited (RFCL), New Delhi, India, and ethanol (Changshu Yangyuan Chemical, China) was used.

Instrumentation and liquid chromatographic conditions

HPLC method was performed using a Waters 2695 quaternary gradient pumps, Waters 2489 Dual λ absorbance detector (Waters, USA). Separation was achieved with ethanol: n-Hexane (0.1 : 50) as eluent at flow rate 0.5 ml/min. The column temperature was maintained at 25°C. Ultraviolet detection was performed at 215 nm. The mobile phase was used as diluent. The run times were used 105 minute and 65 minute for sample and reference standard, respectively, because of the fact that a peak at retention time of 95 minutes was appearing in sample.

Optimization of mobile phase and selection of column

During method development trials, three different chiral columns have been used-Daicel chiralpak AD-H (250 x 4.6 mm), Daicel chiralpak OD-H (250 x 4.6 mm), and Daicel chiralpak AS-H (250 x 4.6 mm). These columns were tried with different mobile phases consisting 2.0 ml and 4.0 ml of ethanol in 1 000 ml of n-hexane. Mobile phase itself was used as the diluent for preparation of reference standards and sample.

Preparation of standard and sample solutions for assay

Preparation of reference solutions (a) and (b)

About 1 00.0 mg of SGB standard was accurately weighed into 20-ml volumetric flask. 10 to 15 ml of diluent was added and Elma Elmasonic S 300 H (Germany) sonicator was used for sonication to dissolve. Volume was made up with diluent and mixed well. 5.0 ml of this solution was diluted to 50 ml with diluent and mixed well (reference solution a). 2.0 ml of reference solution (a) was diluted to 20 ml with diluent and mixed (reference solution b).

Preparation of Reference solution (c)

200.0 mg of RGB sample was accurately weighed into a 20-ml volumetric flask. 10 to 15 ml of diluent was used to dissolve by sonication. 2.0 ml of reference solution (a) was added and volume was made up with diluent.

Preparation of test solution

200.0 mg of sample was accurately weighed into a 20-ml volumetric flask. They were dissolved in diluent by sonication and made up the volume with diluent.

System suitability

The relative standard deviation (RSD) of six replicate injections of reference solution (b) was not more than 5% and resolution between the main peak and SGB peak of reference solution (c) was more than 1.0 set as the system suitability criteria.

Method validation

Method validation is closely related to method development. When a new method is being developed, some parameters are already being evaluated during the "development stage," while in fact, this forms part of the "validation stage." The ICH guidelines achieved a great deal in harmonizing the definitions of required validation parameters, their calculation and interpretation. The international conference on the Harmonization of the Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) has harmonized the requirements in two guidelines. [2],[12] The first one summarizes and defines the validation characteristics needed for various types of test procedures and the second one extends the previous test to include the experimental data required and some statistical interpretation.

Specificity

The ability of the method to determine accurately and specifically the analyte of interest in the presence of other components in a sample matrix (that may be expected to be present in the sample matrix) under the stated conditions of the test (specificity = 100% selectivity).

Solution stability

Drug stability in Active Pharmaceutical Ingredient is a function of storage conditions and chemical properties of the drug and its impurities. Conditions used in stability experiments should reflect situations likely to be encountered during actual sample handling and analysis. Stability data are required to show that the concentration and purity of analyte in the sample at the time of analysis corresponds to the concentration and purity of analyte at the time of sampling.

Linearity

The ability of the method to obtain test results proportional to the concentration of analyte (within a given range). It was evaluated by linear regression analysis, which was calculated by the least square regression method. [13]

Linear range

The range of concentrations or amounts of analyte over which the method gives test results proportional to the concentration of analyte or a linear calibration model can be applied with a known confidence level.

Accuracy

The closeness of agreement between a test result and the accepted reference value. Accuracy of the method was determined by recovery study.

Limit of detection

The limit of detection (LOD) is the point at which a measured value is longer than the uncertainty associated with it. It is the lowest concentration of analyte in a sample that can be detected but not necessary quantified.

LOD = 3.3 × SO / b

Where, SO and b are standard deviations and slope of the calibration line, respectively.

Limit of quantitation

The lowest concentration or amount of analyte that can be determined quantitatively within an acceptable level of repeatability precision and trueness.

Limit of quantitation (LOQ) = 10.0 × SO/b

Ruggedness

The (intra-laboratory tested) behavior of an analytical process when small changes in environment and/or operating condition are made.

Robustness

A measure of the capacity of the analytical procedure to remain unaffected by small but deliberate variations in method-performance parameters, which provides an indication of its reliability during normal usage.


   Results and Discussions Top


Optimization of mobile phase and selection of column

Two different compositions of mobile phases were tried out in three different columns and it was observed that 2.0 ml ethanol in 1 000 ml n-hexane on Chiralpak AD-H produced best resolution and peak shape.

System suitability

Resolution between the main peak and SGB peak of reference solution (c) was found to be 1.19 [Figure 2].
Figure 2: Chromatogram of SGB spiked into RGB

Click here to view


Method validation

Specificity

Specificity of the method was evidenced by comparing blank, SGB, and SGB spiked into RGB solutions [Figure 2], [Figure 3], [Figure 4]. From the experimental data, there are no interfering peaks at the retention times of RGB and SGB from the chromatogram [Figure 2]. Good separation is also seen between RGB and SGB from the spiked chromatogram.
Figure 3: Chromatogram of reference solution (b) of SGB

Click here to view
Figure 4: Chromatogram of RGB of sample

Click here to view


Solution stability

The solution stability was ascertained from HPLC peak area of reference standard samples. The standard sample solutions were kept at room temperature for 48 hours; it was observed that there was no change in peak area of these solutions.

Linearity

Under the experimental conditions, the peak area-concentration plot for the proposed method was found to be rectilinear over the range of 50 to 150% of the specified limit with a correlation coefficient of 0.999 [Figure 5].
Figure 5: Linearity graph of SGB

Click here to view


Accuracy

Analytical method may be considered validated in terms of accuracy if the mean value is within ±10% of the actual value. Recovery of SGB was found in the range of 91.83 to 95.11% [Table 1], which was well within the acceptance criteria.
Table 1: Validation parameters studied for SGB content

Click here to view


Limit of detection and Limit of quantitation

The Limit of detection of SGB was found to be 0.23 microgram / milliliter (μg/ml). The Limit of quantification of SGB was found to be 0.69 microgram / milliliter (μg/ml)

Ruggedness

The ruggedness of the method was evaluated by estimating % RSD of reference solution (b) by two different analysts using different HPLC columns on different days. The average % RSD was found to be 1.27%.

Robustness

Robustness of the method was determined by analyzing the sample with deliberate change in the parameters like (a) flow rate of mobile phase (± 0.1 ml/min), (b) column temperature (± 5°C), (c) wavelength of detection (± 2 nm), and (d) mobile phase composition (± 0.5 ml of ethanol).

(a) Change in flow rate

It was observed that at a flow rate of 0.4 ml/min and 0.6 ml/min, resolution between the main peak and SGB peak of reference solution (c) were found 1.10 and 2.57, respectively. The % RSD values were within desired limits.

(b) Change in column temperature

It was observed that resolutions between RGB and SGB in reference solution (c) were shifted from 1.06 to 1.52 when column temperature changed from 20°C to 30°C.

(c) Change in wavelength of detection

It was observed that changing the wavelength from 213 to 217 nm did not change the % RSD values significantly when compared with injections under normal condition. The resolution main peak and SGB peak of reference solution (c) were found to be 1.07 and 1.15, respectively.

(d) Mobile phase composition

It was observed that resolutions between RGB and SGB in reference solution (c) were shifted from 1.10 to 1.29 when volume of ethanol increased from 1.5 to 2.5 ml with 1 000 ml n-Hexane and the % RSD values were within desired limits.


   Conclusions Top


The proposed method does not require any laborious clean up procedure before measurement. The validation data demonstrate good accuracy, which proves the reliability of the proposed method. Hence, the validated method can be used for routine determination of RGB present in SGB in quality control laboratories in the pharmaceutical industry and thus commercial viability for synthesis of Linezolid has been enhanced.


   Acknowledgement Top


Thanks to Mr. Glenn Saldanha, Chairman & MD, Glenmark Pharmaceuticals Limited and Dr. Milind Gharpure Senior Vice President - Process Research Glenmark Generics Limited for their consistent support and the management of the Glenmark Generics Limited.

 
   References Top

1.Snyder LR, Kirkland JJ, Glajch JL. Practical HPLC Method Development. New York: Wiley-Interscience; 1997.  Back to cited text no. 1
    
2.ICH Steering Committee, International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use, Validation of Analytical Procedures: Methodology, November 6, Geneva, Switzerland; 1996.  Back to cited text no. 2
    
3.Kloosterman M, Elferink VH, Vaniersel J, Roskam JH, Meijer EM, Hulshof LA. Lipases in the preparation of b-blockers. Trends Biotechnol 1988;6:251-6.  Back to cited text no. 3
    
4.Kasai N, Suzuki T, Furukawa Y. Chiral C3 epoxides and halohydrins: their preparation and synthetic application. J Mol Catal B Enzym 1998;4:237-52.  Back to cited text no. 4
    
5. Palomo JM, Segura RL, Mateo C, Terreni M, Guisan JM, Fernandez- Lafuente R. Synthesis of enantiomerically pure glycidol via a fully enantioselective lipase-catalyzed resolution. Tetrahedron Asymmetry 2005;16:869-74.  Back to cited text no. 5
    
6.Brickner SJ, Hutchinson DK, Barbachyn MR, Manninen PR, Ulanowicz DA, Garmon SA, et al. Synthesis and antibacterial activity of U-100592 and U-100766, two oxazolidinone antibacterial agents for the potential treatment of multidrug-resistant gram-positive bacterial infections. J Med Chem 1996;39:673-9.  Back to cited text no. 6
[PUBMED]  [FULLTEXT]  
7.Weidner-Wells MA, Boggs CM, Foleno BD, Melton J, Bush K, Goldschmidt RM, et al. Novel piperidinyloxy oxazolidinone antibacterial agents. Diversification of the N-Substituent. Bioorgan Med Chem 2002;10:2345-51.  Back to cited text no. 7
    
8.Matsuda M, Endo Y, Fushiya S, Endo T, Nozoe S. Cytotoxic 6-substituted 5,6-dihydro-2H-pyran-2-ones from a brazilian medicinal plant, chorisiacrispiflora. Heterocycles 1994;38:1229-32.  Back to cited text no. 8
    
9.Waechter AI, Ferreira ME, Fournet A, Rojas de Arias A, Nakayama H, Torres S, et al. Experimental treatment of cutaneous leishmaniasis with argentilactone isolated from Annona haematantha. Planta Med 1997;63:433-5.  Back to cited text no. 9
[PUBMED]  [FULLTEXT]  
10.Ahuja S, Rasmussen H. HPLC method development for pharmaceuticals. Elsevier Sciverse Science Direct. 2007. Available from: http://www.sciencedirect.com/science/publication?issn=01496395and volume=8. [Last accessed on 2011 May 30].  Back to cited text no. 10
    
11.Subramanian G. Chiral separation techniques: a practical approach. Weinheim: Wiley-VCH Verlag Gmbh and Co; 2007. p. 1-592.  Back to cited text no. 11
    
12.International Conference on Harmonization: European Union, Japan, USA, Validation of Analytical procedures: Text and Methodology, 1994. Published in FDA federal register, 1995;60:11260.  Back to cited text no. 12
    
13.Christian GD. Analytical Chemistry. Singapore: John Wiley and Sons PTE. LTD; 2004. p. 102-3.  Back to cited text no. 13
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1]


This article has been cited by
1 A rapid and validated reverse phase liquid chromatographic method for determination of imiquimod from topical cream formulations
Madhusudan T. Bachute,Shankar L. Turwale
Journal of Pharmacy Research. 2013; 6(1): 73
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Experimental
    Results and Disc...
   Conclusions
   Acknowledgement
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2255    
    Printed165    
    Emailed0    
    PDF Downloaded302    
    Comments [Add]    
    Cited by others 1    

Recommend this journal